«Anesteziologiya i Reanimatologiya» №3 2013


ASSESSMENT OF FUNCTIONAL RESIDUAL CAPACITY AND OXYGEN CONSUMPTION DURING PEEP



Yüklə 80,46 Kb.
Pdf görüntüsü
səhifə11/14
tarix26.12.2016
ölçüsü80,46 Kb.
#3494
1   ...   6   7   8   9   10   11   12   13   14
ASSESSMENT OF FUNCTIONAL RESIDUAL CAPACITY AND OXYGEN CONSUMPTION DURING PEEP 

TRIAL PROCEDURE  

Ruchina E.V., Charnin S.A., Lebedinsky K.M., Mazurok V.A.  

North-Western State Medical University named after I.I. Mechnikov, 191015, Saint-Petersburg, Russia  

Purpose of the study was to compare the functional residual capacity and oxygen consumption with 

compliance and partial pressure of oxygen during automatic PEEP trial procedure —  "PEEP INview". 

Material and methods; Dynamics of functional residual capacity, oxygen consumption, compliance and 

partial pressure of oxygen were studied in 14 patients at cardiac surgical unit during automatic PEEP trial 

procedure  —  "PEEP INview". Results; Signifi cant dynamics of compliance and carbon dioxide production 

was not found. Partial pressure of oxygen cannot be used as the guideline in this study. Functional residual 

capacity increased in each incremental level of PEEP. Oxygen consumption increased during the PEEP 

changing from 6 to 8 cmH2O. Conclusion; It is defined that oxygen consumption is sensitive to PEEP  level. 

Therefore the maximum diagnostics value of functional residual capacity measuring is expected in patients 

with restrictive lung disease.  

Key words: mechanical ventilation, functional residual capacity, oxygen consumption, compliance, positive 

end-expiratory pressure  

 

AN-1303-054 