BRAEBON Medical Corporation

                                                                                                D.0570.2..40600.1



Legal Notices

Copyright ©2008-2010 by BRAEBON™ Medical Corporation. All rights reserved. MediPalm is a trademark of 

BRAEBON Medical Corporation. Alice

®

5 is a registered trademark of Philips Respironics. 



 

BRAEBON™ is an ISO 13485 compliant company.

 

Technical Support: 1-888-462-4841 (9 am - 5 pm EST) 



E–mail:support@braebon.com 

Customer Sales or Service: 1-888-462-4841 

 E–mail:info@braebon.com



www.braebon.com

1-100 Schneider Road

Kanata, Ontario, 

K2K 1Y2


102-2981 Ford St. Extension, 

Ogdensburg, NY. 

13669-3474


Q-RIP Respiratory Effort System

1

Intended Use



The BRAEBON Medical Corporation Q-RIP™ Respiratory Effort System™ is intended for use during sleep 

disorder studies as a measure of respiratory effort for recording by a data acquisition system. The sensor uses 

respiratory inductance plethysmography (RIP) technology and plugs directly into an electrode headbox or a 

DC amplifier.

The target population of the Q-RIP Respiratory Effort System is all children and adult patients who are 

screened during sleep disorder studies. The screenings occur in a sleep laboratory or other clinical setting.

The Q-RIP Respiratory Effort System intended to be used only by or on the order of a physician.

The Q-RIP is for diagnostic use only; it is NOT intended as an apnea monitor, and it is NOT to be used in 



life sustaining situations. 

U.S. Federal law restricts this device to sale by or on the order of a physician.



Do NOT steam autoclave or gas sterilize the Q-RIP interface or belts as damage will result.

Contents of the Q-RIP Kit, 0570

Part Number

Description

Quantity

0571


Q-RIP Interface

1

0572



Q-RIP Effort Belt, adjustable from 24” to 48”

2

0573



Q-RIP Belt to Controller Cable

2

0574



Q-RIP Controller Output Cable

3

820-0005-01



1.5 V AA battery

2

User Guide



1

Multi-Language User Guide

1

Warnings


Note:  

Unlike piezo belts, RIP belts must be put on tightly. You can 

easily loosen the belt once it is on. It is more difficult to tighten the RIP 

belt once it is on the body.

 

Q-RIP Respiratory Effort System



Description

The Q-RIP System consists of two respiratory effort belts (0572) and one Q-RIP interface (0571). The two 

respiratory effort belts are worn by the patient around the waist and abdomen to measure respiratory effort. 

The belts are plugged into the Input ports on the interface using the cables provided with the belts. Three 

output cables plug into the Output ports on the interface and into the PSG system. These output cables send 

the abdominal, chest and sum signals to the PSG system. 



Figure 1: Q-RIP System

Q-RIP Respiratory Effort System

3

To fit the chest belt



1. Tighten the belt as much as possible; it is much 

easier to loosen a tight belt on the body than to 

tighten a loose belt.

2. Wrap the chest belt around the chest and snap it 

closed. It is not necessary to place the belt on bare 

skin; we recommend you place the belts over 

bedtime clothing.

3. Ensure the belt is located above or below the nipple 

line and loosen for a firm but comfortable fit.

4. Use medical tape to secure the belt to clothing to 

prevent shifting.

5. Take one of the Y-shaped cables and plug one safety 

pin connector into the left socket on the belt and the 

other safety pin connector into the right socket. It 

does not matter which colored pin goes into the 

socket.


6. Plug the keyhole connector into the keyhole input of 

the Q-RIP interface, labelled 



CHEST

.

To fit the abdominal belt



1. As with the chest belt, tighten the belt as much as 

possible before putting it on. 

2. Wrap the belt around the belly and snap closed. 

Loosen for a firm but comfortable fit.

3. Use medical tape to secure the belt to clothing to 

prevent shifting.

4. Take the remaining Y-shaped cable and plug one 

safety pin connector into the left socket located on 

the belt and the other safety pin connector into the 

right socket.

5. Plug the keyhole connector into the keyhole input on 

the Q-RIP interface labelled 

ABDOMEN

.

Figure 1: Chest belt plugged into the Q-RIP 



interface

Figure 2: Chest and abdominal belt plugged 

into the Q-RIP interface

 

Q-RIP Respiratory Effort System



Output to PSG

1. Connect the three output cables to the three output 

ports on the Q-RIP.

2. Plug the other ends into your PSG or headbox.

System Settings

Refer to your sleep system user guide for the manufacturers’s recommended settings. General recording 

system settings:

Low Frequency Filter: 0.1 Hz or lower (TC=1 sec. or longer)



High Frequency Filter: 5 Hz to 15 Hz

50/60 Hz Notch Filter: optional



Sensitivity/Gain 2 mV/cm or 4 000 gain (increase or decrease as needed).

Starting the Q-RIP

1. Set the Gain on the front of the Q-RIP Interface, refer to Table 1 on page 7 for the gain ranges.

2. Turn the power ON.

3. Wait two minutes for the signals to settle. This means that when you review the data, the signals in the 

first two minutes may not look correct. 

Stopping the Q-RIP

1. At the end of the study, turn the Q-RIP power OFF. If you forget, the system will automatically turn off 

after 14 hours.



Figure 3: Q-RIP interface with inputs and 

outputs

Q-RIP Respiratory Effort System

5

Signals



The Q-RIP collects the chest, abdominal and sum signal. The sum signal is useful as an aid for detecting 

paradoxical breathing. When the chest and abdominal signals are completely out of synchronization, the sum 

channel will tend toward flatness.

Figure 4: Sample normal and paradoxical breathing


 

Q-RIP Respiratory Effort System



Troubleshooting

If you have difficulty using this product, please verify the following:

If you still experience trouble with the product, contact technical support by telephone at 1.888.462.4841 x 

214 or by E-mail at support@braebon.com. Visit our website for solutions to Frequently Asked Questions 

(FAQs): 


www.braebon.com/FAQ

Maintaining the Q-RIP



The Q-RIP Respiratory Effort System is a non-sterile device. It is classified as a non-critical item because it 

only contacts clothing and intact skin.

Cleaning the Q-RIP Interface

:

1. Moisten a clean cloth with water.



2. Gently, wipe the Q-RIP interface with the damp cloth.

3. Ensure that the Q-RIP interface is completely dry before using. 

For additional information, refer to the Association for Professionals in Infection Control and Epidemiology 

(APIC) Guidelines for selection and use of disinfectants (American Journal of Infection Control. Vol. 18, 

No. 2, April 1990).

Cleaning the Q-RIP Belts

Place RIP belts into a mesh laundry bag to prevent tangling. Machine-wash on the gentle cycle, hang to dry.

Problem

Solution

No signal

If you have just turned the unit on, wait 2 minutes



Check batteries

Check belt connections



Check connections to PSG

Check the gain setting is in the correct position



Noisy signal

Belts not tight enough.



Check the gain setting is in the correct position

Q-RIP was 

submerged in liquid

Contact BRAEBON



 Medical Corporation at 1.888.462.4841 x 214. Please 

have the product model and serial number available when you call.

Caution:  Sterilization of the Q-RIP is NOT required. Do NOT steam autoclave any 

component of the Q-RIP System as damage will occur and void the warranty. 

Caution:  Do NOT immerse the Q-RIP Interface in any liquids because damage 

will result. Water damage will void the warranty.


Q-RIP Respiratory Effort System

7

Table 1: Specifications



Dimensions

5.0 x 2.8 x 0.9 inches (125 x 70 x 24 mm)



Weight

2.9 oz. with batteries (82 g)



Case

Plastic (ABS)



Power

2 x AA Alkaline batteries; 2 x NiMH or 2 x NiCD rechargeable 

batteries may be used. Do not mix different types of batteries; do 

not mix old and new batteries.



Battery life

Up to 70 nights



Indicators

One bi-colored (red/green) status LED. Green blink every 10 

seconds indicates unit operating; red blink every two seconds 

indicates a bad battery.



Gain Switch

Set the Gain to DC, AC High or AC Low



DC

Set the gain for a DC amplifier

0 V - 1 V

AC High

Set the gain for Alice 5, MediPalm

± 250 mV

AC Low

Set the gain for most headboxes, PSG systems

 ± 5 mV

Table 2: Q-RIP Belts



Part # 

Description

0572

Q-RIP Effort Belt, medium size, adjustable from 24”-48'.



0572L

Q-RIP Effort Belt, large size, adjustable from 48”-72”.



0572Ped

Q-RIP Effort Belt, small size, adjustable from 12”-24”.



 

Q-RIP Respiratory Effort System



Table 3: Accessories

Part #

Description

Connector

0352

36” (90 cm) Dual 1 mm keyhole connector to 1 mm keyhole 

connector

0353

18” (45 cm) Dual 1 mm (keyhole) connector to 1/8” (3.5 

mm) male phone connector

0354

18” (45 cm) Dual 1 mm (keyhole) connector to 1/8” (3.5 

mm) female phone connector

0355

18” (45 cm) Dual 1 mm (keyhole) connector to 3/32” (2.5 

mm) male phone connector

0356

18” (45 cm) Dual 1 mm (keyhole) connector to 3/32” (2.5 

mm) female phone connector

0357

18” (45 cm) Dual 1 mm (keyhole) connector to RJ11 

connector


Q-RIP Respiratory Effort System

9

0358

18” (45 cm) Dual 1 mm (keyhole) connector to RJ10 

connector



0573

Q-RIP Belt to Controller Cable: 84” Dual 1 mm (keyhole) 

connector to two 1.5 mm touchproof connectors. 

0574

Q-RIP Controller Output Cable: 18” (45 cm) Dual 1 mm 

(keyhole) connector to two 1.5 mm touchproof connectors

Table 4: Environmental Conditions

Operating Conditions

Storage Conditions

Temperature 

(in degrees Celsius)

10

C–40C



5

C–50C


Relative Humidity

10%–85% 


(Non-condensing)

10%–85%


Table 3: Accessories

Part #

Description

Connector

10 

 

Q-RIP Respiratory Effort System



Repairing the Q-RIP

The Q-RIP does not contain serviceable parts.

The Q-RIP is only repairable by BRAEBON

– trained personnel. Opening of the Q-RIP interface by 



unauthorized individuals will void both the safety of the Q-RIP and the terms and conditions of the Q-RIP 

warranty.

If your Q-RIP or any of its accessories are damaged, please contact BRAEBON

 immediately at 



1.888.462.4841 x 214.

Battery Disposal

Please dispose of used batteries responsibly. To locate a battery disposal sight near you, go to 

www.ehso.com.

Warning:  Unauthorized opening of the Q-RIP will void both the safety of the Q-RIP 

and the terms and conditions of the Q-RIP warranty.

Do NOT use or attempt to service damaged parts. 


Q-RIP Respiratory Effort System

11

Warranty



BRAEBON

 

MEDICAL



 

CORPORATION

 warrants to the first consumer that this Ultima Q-RIP Sensor, when shipped in its original container, 

will be free from defective workmanship, performance and materials and agrees that it will, at its option, either repair the defect or 

replace the defective Q-RIP or part thereof at no charge to the purchaser for parts or labor for a time period of one year from the date 

of purchase. The warranty described herein shall be the sole and exclusive warranty granted by 

BRAEBON

 

MEDICAL



 

CORPORATION

 and 

shall be the sole and exclusive remedy available to the purchaser. Use of the Q-RIP constitutes total and complete acceptance of 



this warranty. Correction of defects, in the manner and for the time period described herein, shall constitute complete fulfillment of all 

liabilities and responsibilities of 

BRAEBON

 

MEDICAL



 

CORPORATION

 to the purchaser with respect to the Q-RIP and shall constitute full 

satisfaction of all claims, whether based on contract, negligence, strict liability or otherwise. In no event shall 

BRAEBON

 

MEDICAL



 

CORPORATION

 be liable, or in any way responsible, for any loss of revenues or damage, direct, incidental, or consequential, including 

property damage, loss of profit, or personal injury resulting from the use or misuse of, or the inability to use this product. Nor shall 

BRAEBON

 

MEDICAL



 

CORPORATION

 be liable, or in any way responsible, for any damages or defects in the Q-RIP which were caused by 

abuse, misuse, tampering, neglect, incorrect battery type, or repairs or attempted repairs performed by anyone other than an 

authorized service person. This warranty covers the Q-RIP Model 0571 only. Accessories and consumables have separate 

warranties with different coverage periods.



Warning:  Unauthorized opening of the Q-RIP will void both the safety of the Q-RIP 

and the terms and conditions of the Q-RIP warranty.

Caution:  Failure to use the correct battery type as stated in this User Guide will void 

the warranty. 

12 

 

Q-RIP Respiratory Effort System



Notes

BRAEBON Medical Corporation

                                                                                                D.0570.2..406001



In Canada:

1-100 Schneider Road

Kanata, Ontario, 

K2K 1Y2


In the US: 

102-2981 Ford St. Extension, 

Ogdensburg, NY

13669-3474

Customer Sales, Technical Support: 1-888-462-4841 (9 am - 5 pm EST) 