The measurement uncertainty in the calibration of a thermometer depends on the calibration method used, the uncertainty contri



Yüklə 187,5 Kb.
səhifə4/12
tarix26.12.2016
ölçüsü187,5 Kb.
#3354
1   2   3   4   5   6   7   8   9   ...   12
uTra: repeatability of measurements during a calibration. We assume at least 50 measurements of reference standard was made at each calibration point. Standard deviation of mean is calculated.

Example: uTra= 0.002°C


  • uTrcon: uncertainty contribution due to a possible heat conduction by the reference standard. Tests should be made at different immersion depths.

Example: pulling the reference standard 20 mm out of the bath led to a temperature change of 2 mK (which due to the temperature variations of the bath could be estimated only inaccurately).


1   2   3   4   5   6   7   8   9   ...   12




Verilənlər bazası müəlliflik hüququ ilə müdafiə olunur ©azkurs.org 2024
rəhbərliyinə müraciət

gir | qeydiyyatdan keç
    Ana səhifə


yükləyin