uTra: repeatability of measurements during a calibration. We assume at least 50 measurements of reference standard was made at each calibration point. Standard deviation of mean is calculated.
Example: uTra= 0.002°C
uTrcon: uncertainty contribution due to a possible heat conduction by the reference standard. Tests should be made at different immersion depths.
Example: pulling the reference standard 20 mm out of the bath led to a temperature change of 2 mK (which due to the temperature variations of the bath could be estimated only inaccurately).