61
shifted compared to pure polystyrene what may be caused by the presence of scintillating
dopants. However, the range of o-Ps lifetime is similar to the results described in
article [97].
Other structural changes, visible on the diagram
of o-Ps intensity I
3
(bottom panel
of Fig. 29), in 280 K and 260 K may be also correlated with structural changes in the
doped polymer. The probable reason of the changes is crystalline - like organization of
molecules in some regions of the amorphous polymer. One can find in the literature [98]
information about phase transitions in polystyrene at low temperatures which is so-called
beta-transition (between 283 K and 333 K). This point can be identified with the phase
transition point found in our results in J-PET polystyrene - based scintillator (in 280 K).
Red points in Fig. 24 denote rapid changes of I
3
dependence on time when the sample was
stored in particular temperature. The measurement confirmed
that o-Ps production is
unstable in time (intensity I
3
decreases with time).
Differences between results obtained in our experiment and the experiment
described in [97] come from the presence of dopants. It is known that even small amount
of impurities may significantly change temperatures of particular phase transitions. For
example in polypropylene copolymers and blends described in article [99] one can observe
the shift of glass transition in comparison to sample of pure polystyrene. Because of the
presence of admixtures, several regions of crystalline-like organization may be formed. It
results in additional structural transitions. Transition denoted as T
β
is probably connected
with the presence of dopants.
Considering
the transition T
γ
, it is clearly visible in the I
3
diagram but almost
invisible in diagram of τ
3
. This indicates that the transition is correlated with energetic
changes
in the molecule, not a geometrical reconfiguration.
It was mentioned, that I
3
of o-Ps can be influenced by the thermal history of the
sample. It is visible in the hysteresis observed for heated and cooled down scintillator.
Because of that, an additional measurements of thermal stability were carried out. PAL
spectra
were collected for a long time, equal to at least 15 hours, in three temperatures:
123 K, 173 K and 298 K. Results of the measurements are shown in Fig. 30.