Appendix A
Time-elapsed plots
The plot graph below is a representative example of continuous sensor tracing vs. reference blood glucose
reading, where sensor showed excellent performance. The open circles (o) on the graph represent the
meter calibration readings. The closed circles (•) represent the reference blood glucose readings, and the
solid line (—) represents the sensor glucose value.
Glucose Conc
(mg/dL)
Day 1
Day 4
Day 2
Day 3
Appendix A
220
The plot graph below is a representative example of continuous sensor tracing vs. reference blood glucose
reading, where sensor showed average (typical) performance. The open circles (o) on the graph represent
the meter calibration readings. The closed circles (•) represent the reference blood glucose readings, and
the solid line (—) represents the sensor glucose value.
Glucose Conc
(mg/dL)
Day 1
Day 4
Day 2
Day 3
Sensor accuracy
221
Appendix A
The plot graph below is a representative example of continuous sensor tracing vs. reference blood glucose
reading, where sensor showed poor performance. The open circles (o) on the graph represent the meter
calibration readings. The closed circles (•) represent the reference blood glucose readings, and the solid
line (—) represents the sensor glucose value.
Glucose Conc
(mg/dL)
Day 1
Day 4
Day 2
Day 3
Appendix A
222
Performance results in children and adolescents
The performance of the Guardian RT in Children and Adolescents was evaluated in a clinical study.
Guardian RT results were compared to capillary blood glucose measured by a LifeScan OneTouch
®
Ultra
®
Meter. Sixty subjects with Type I diabetes participated in a three-site out-patient study. Subjects ranged in
age from 7 to 17 years old. Each subject wore a Guardian RT system, which records a sensor glucose value
every five minutes. Subjects were instructed to perform at least seven blood glucose meter measurements
each day.
Accuracy of Guardian RT readings
In this study, each blood glucose meter measurement was paired with the corresponding Guardian RT
reading (taken every 5 minutes). Pairing was done by selecting the Guardian RT value closest in time to
the blood glucose meter result. Agreement was analyzed by comparing paired glucose measurements.
Agreement between the matched pair was estimated by evaluating the difference between the Guardian
RT reading and the blood glucose meter measurement. The difference between them was calculated as a
percentage of the blood glucose meter measurement (Mean Absolute Percent Difference). The bias was
also calculated, and it is defined as the overall difference between the Guardian RT glucose values and the
blood glucose meter values. The paired glucose measurements are summarized in the following table
Number of Paired Glucose Measurements
2599
Mean Absolute Percent Difference (± SD)
19.0 ± 19.7%
Bias
-6.0 mg/dL (-0.3 mmol/L)
Sensor accuracy
223
Appendix A
The accuracy of the Guardian RT was also evaluated by calculating the percentage of Guardian RT readings
within 20% and within 30% of the blood glucose meter reading (or within 20 mg/dL (1.1 mmol/L) in the low
glucose range). Results are shown in the following table.
Plasma Glucose
Range (mg/dL)
Plasma Glucose
Range (mmol/L)
Number of Paired
Readings
Percent Within
20%
Percent Within
30%
Overall
2599
68%
82%
40-80*
2.2-4.4
360
51%
----
> 80-120
> 4.4-6.7
482
60%
77%
> 120-240
> 6.7-13.3
1055
74%
90%
> 240
> 13.3
702
75%
89%
*For the Low glucose range, 40-80 mg/dL (2.2-4.4 mmol/L), the value shown is the percent within 20
mg/dL (1.1 mmol/L).
The Clarke Error Grid was used to assess the clinical relevance of the differences between the Guardian RT
readings and the comparative blood glucose meter measurements. The Clarke Error Grid divides a
correlation plot into 5 zones. See the following table.
Results in zones A and B are considered clinically acceptable, while results in zones C, D, and E are
potentially dangerous and, therefore, clinically significant errors. The Clarke Error Grid zones are labeled
on the correlation plot.
Zone
Description
A
Clinically accurate, would lead to correct treatment decisions
B
Would lead to benign decisions or no treatment
C
Would lead to over-correction of normal glucose levels
D
Would lead to failure to detect and treat high or low glucose levels
E
Would lead to erroneous treatment decisions
Appendix A
224
The following figure is a correlation plot of Guardian RT readings versus readings from the reference
method, the blood glucose meter. It is overlaid with the Clarke Error Grid. The total number of paired data
points is 2599.
Meter BG (mg/dL)
Sensor BG (mg/dL)
A (1733, 66.7%)
B (665, 25.6%)
C (7, 0.3%)
D (193, 7.4%)
E (1, 0.0%)
The percent of Guardian RT readings in the above graph are presented in the table below according to the
percentage of points falling within each zone (A-E). Results are further broken down (stratified) according
to the range of glucose concentrations.
Sensor accuracy
225
Appendix A
Stratified Clarke Error Grid Analysis
Range of
Comparative
Glucose
Readings
(mg/dL)
Total
Count
A + B
A
B
C
D
E
40-80
360 (13.9%)
201
(55.8%)
141
(39.2%)
60 (16.7%)
1 (0.3%) 157
(43.6%)
1 (0.3%)
81-120
482 (18.5%)
478
(99.2%)
287
(59.5%)
191
(39.6%)
4 (0.8%) 0 (0%)
0 (0%)
121-240
1055
(40.6%)
1053
(99.8%)
782
(74.1%)
271
(25.7%)
2 (0.2%) 0 (0%)
0 (0%)
240-400
702 (27.0%)
666
(94.9%)
523
(74.5%)
143
(20.4%)
0 (0%)
36 (5.1%)
0 (0%)
Overall
2599
(100.0%)
2398
(92.3%)
1733
(66.7%)
665
(25.6%)
7 (0.3%) 193 (7.4%)
1 (0.0%)
Appendix A
226
Low and High Alerts in children and adolescents
The ability of the Guardian RT to detect high and low glucose levels was measured in the same clinical
study. Since it is important to set the alert levels in a conservative fashion, the Low Glucose Alert should
be set at a value slightly higher than the value of blood glucose you want to detect, and the High Glucose
Alert should be set at a value slightly lower than the value of blood glucose you want to detect.
NOTE:
Please ask your healthcare professional which low and high alert setting is best for you.
The Low Glucose Alert
The Low Glucose Alert was evaluated for its ability to detect glucose levels at 70 mg/dL (3.9 mmol/L), or
below, using the blood glucose meter. As a reference, with the Low Glucose Alert set at 70 mg/dL (3.8
mmol/L), 24% (59/244) of low glucose events were detected by the Guardian RT. Better detection of low
blood glucose can be obtained by setting the Low Glucose Alert level higher. For example, setting the Low
Glucose Alert at 90 mg/dL (5.0 mmol/L), instead of 70 mg/dL (3.9 mmol/L), increases the ability to
detect low blood glucose levels from 24% to 70% (see the following table).
Sometimes the Guardian RT will alert when the blood glucose levels are not low. When the Guardian RT
Low Alert was set at 70 mg/dL (3.9 mmol/L) in this study, 48% of the results were considered false alerts
(actual blood glucose values are greater than 85 mg/dL (4.7 mmol/L)). This percentage may be
exaggerated because blood glucose may be dropping when the Guardian RT alerts.
The table below shows the percent of Low Glucose readings correctly identified by the Guardian RT for
specific settings.
Guardian RT Low Alert
Setting (mg/dL)
Guardian RT Low Alert
Setting (mmol/L)
True Alert Rate*
False Alert Rate**
70
3.9
24%
48%
80
4.4
52%
46%
90
5.0
70%
52%
100
5.6
85%
57%
* True Alert Rates are the % of times when the glucose level was at or below the alert setting and the alert
sounded.
** False Alerts Rates are the % of times when the Guardian RT Sensor alarmed but the blood glucose level
was greater than the alert setting.
Sensor accuracy
227
Appendix A
Increasing the Low Alert settings will improve the ability to detect low blood glucose events, but it will
also increase the frequency of Guardian RT false alerts for blood glucose levels not below the target value.
You should consider this trade-off between the improved ability to detect true low blood glucose versus
the increased number of false alerts when setting the low alert threshold.
The High Glucose Alert
The High Glucose Alert was evaluated for its ability to detect glucose levels at 250 mg/dL (13.8 mmol/L),
or above, using the blood glucose meter. As a reference, with the High Glucose Alert set at 250 mg/dL
(13.8 mmol/L), 64% (404/632) of high glucose events were detected by the Guardian RT. Better detection
of high blood glucose can be obtained by setting the High Glucose Alert level lower. For example, setting
the High Glucose Alert at 190 mg/dL (10.6 mmol/L), instead of 250 mg/dL (13.8 mmol/L), increases the
ability to detect high blood glucose levels from 64% to 94% (see the following table).
Sometimes the Guardian RT will alert when the blood glucose levels are not high. When the Guardian RT
High Alert was set at 250 mg/dL (13.8 mmol/L) in this study, 13.1% of the results were considered false
alerts (actual blood glucose values are less than 225 mg/dL (12.5 mmol/L)). This percentage may be
exaggerated because blood glucose may be rising when the Guardian RT alerts. The following table shows
the percent of High Glucose readings correctly identified by the Guardian RT for specific settings.
Guardian RT High Alert
Setting (mg/dL)
Guardian RT High Alert
Setting (mmol/L)
True Alert Rate*
False Alert Rate**
190
10.6
94%
40%
200
11.1
91%
36%
225
12.5
81%
21%
250
13.8
64%
13%
* True Alert Rates are the % of times when the glucose level was at or above the alert setting and the alert
sounded.
** False Alerts Rates are the % of times when the Guardian RT Sensor alarmed but the blood glucose level
was lower than the alert setting.
Decreasing the High Alert settings will improve the ability to detect high blood glucose events, but it will
also increase the frequency of Guardian RT false alerts for blood glucose levels not above the target value.
You should consider this trade-off between the improved ability to detect true high blood glucose versus
the increased number of false alerts when setting the high alert threshold.
Appendix A
228
Glossary
A
A1C (HbA1c) test - A 2-3 month
average of blood glucose
values expressed in percent.
The normal range varies with
different labs and is
expressed in percent (such as
4 - 6%).
Accept - Pressing the ACT button
to approve the selection or
setting.
Active insulin - Bolus insulin
that has been delivered to
your body, but has not yet
been used.
Adverse reaction - An
unexpected, unpleasant or
dangerous reaction to a
sensor when it is inserted into
the body. An adverse reaction
may be sudden or may
develop over time.
Alarm - Audible or vibrating
(silent) notice that indicates
the pump is in Attention
mode and immediate
attention is required. Alarms
are prefixed in the alarm
history with the letter A.
Alarm clock - Feature you can
set to go off at specified
times of the day.
ALARM HISTORY - Screen that
displays the last 36 alarms/
errors that have occurred on
your pump.
Alarm icon - a solid circle that
shows at the top of the screen
and the pump beeps or
vibrates periodically until the
condition is cleared (see
Attention mode).
Alarm Snooze - Once a METER
BG NOW alert occurs, the
pump will not repeat the alert
until after this selected Alarm
Snooze time.
Alert - Audible or vibrating
(silent) indicator that notifies
you the pump needs attention
soon or that you should be
reminded of something.
Insulin delivery continues as
programmed.
Alert icon - an open circle that
appears at the top of the
screen and the pump beeps or
vibrates periodically to
remind you of the condition
(see Special mode).
Attention mode - Operating
mode that stops all current
insulin delivery. This mode
indicates an alarm or
condition exists that requires
immediate attention.
B
Backlight - Pump screen light.
Turns on whenever the
button is pressed or in
combination with the
button if not on the HOME
screen. The backlight also
appears during alarms and
alerts, except for LOW
BATTERY alert and OFF NO
POWER alarm.
Basal patterns - The user can
program up to three different
basal rate patterns into the
pump for the delivery of basal
insulin: Standard, A, and B.
For each pattern, there is an
option of setting up to 48
basal rates.
Glossary
229
Glossary
Basal rate - The pump setting
that provides a continuous
infusion of insulin to keep the
blood glucose stable between
meals and during the night.
Basal insulin mimics
pancreatic insulin delivery—
which meets all the body's
non-food related insulin
needs.
BASAL REVIEW screen - shows
the basal rates programmed
in the pump, with the 24-hour
total for each rate.
BG - Blood Glucose
BG Reminder - Feature that you
can set to remind you to
check your blood glucose
after a bolus.
BG Target - Normal blood
glucose level
BG Units - Blood glucose units
used by the pump (mg/dL or
mmol/L).
Block - Feature that restricts
access to all programming
except suspend, selftest and
the delivery of a bolus with
the remote control.
Blood glucose (BG) - A form of
digested sugar in the
bloodstream. Glucose is the
major source of energy for
living cells, and is carried to
each cell through the blood
stream. Cells cannot use
glucose without the help of
insulin.
Bolus - A dose of insulin given to
cover an expected rise in
blood glucose (such as the
rise after a meal) or to lower
a high blood glucose down to
target range.
BOLUS HISTORY - This screen
displays the last 24 boluses
delivered by your pump.
Bolus Wizard
®
feature -
Calculates the bolus amount
based on personal information
of the pump user.
C
Cal Reminder - The value of the
Cal Reminder is the amount of
time before the current
calibration value expires
when the user wants to be
reminded to calibrate by
having the pump issue a
METER BG BY alert. For
example, if the Cal Reminder
is set to two hours, the METER
BG BY alert will occur two
hours before the calibration is
required.
Calibrate - Check, adjust, or set
to a standard (calibrate your
pump).
Cannula - A short, thin, and
flexible tubing at the end of
the infusion set that is
inserted into the
subcutaneous tissue to deliver
insulin.
Carb ratio (Carbohydrate ratio)
- Used when counting
carbohydrates in grams. The
amount of carbohydrates
covered by one unit of insulin.
(Also see exch ratio.)
Carb units - The food entry
when using the Bolus Wizard
feature. Entered as
(carbohydrate) grams or
exchanges.
Carbohydrate - One of the three
main sources of energy or
calories in food.
Carbohydrates are mainly
sugars found in fruits, milk,
and starches that the body
breaks down into glucose.
CH - Carbohydrate
Contraindication - A condition
that makes a specific
treatment or procedure NOT
advisable. A contraindication
is literally contra- (against) an
indication, against something
that is indicated as advisable
or necessary.
Correction bolus - The amount
of insulin needed to return a
high blood glucose level back
down to target range.
D
Daily totals - Shows the total
insulin delivered (basal and
bolus) in the last 24-hours.
Maximum records: 14 days
Default - A setting or value for
your pump that is assigned
automatically by your system.
Some default settings cannot
be changed; other default
Glossary
230
settings remain in effect until
you change them.
Diabetic Ketoacidosis (DKA) - A
serious condition that occurs
when the insulin levels are
low, blood glucose level are
elevated, and the body uses
fat for energy. This process
produces ketones which upset
the body's acid-base balance
leading to a potentially life
threatening situation.
DKA - Diabetic Ketoacidosis
Dual Wave
®
bolus - Combination
of a Normal bolus that is
delivered immediately, then
followed by a Square Wave
bolus. The Square Wave
portion is delivered evenly
over a period of time.
Duration - Amount of time it
takes to administer a bolus or
basal delivery. Also, length of
time for an action or
condition.
E
Easy bolus™ - Method of delivery
for a Normal bolus using the
Easy Bolus button
.
Electromagnetic compatibility -
The condition that exists
when systems and devices
that use electromagnetic
energy operate properly
without causing or suffering
accidental electromagnetic
interference to or from other
electromagnetic devices.
Electrostatic discharge - The
rapid, spontaneous transfer of
electrostatic charge induced
by a high electrostatic field.
The charge usually flows
through a spark, known as
static discharge, between two
objects at different
electrostatic levels as they
approach each other (for
example, when people touch
each other after dragging
their feet on a carpet).
Exch ratio - (Exchange ratio)
Used when counting
carbohydrates as exchanges.
The amount of insulin
required to cover one (1)
carbohydrate exchange. (Also
see carb ratio.)
Exchange system - A system of
estimating carbohydrate and
calorie intake that groups by
type. Each exchange group
lists similar types of food in
serving sizes. A person can
exchange, trade, or substitute
a food serving in a particular
group for another food serving
in the same group.
Express bolus - Method of
delivery for any bolus type
using the express bolus button
.
F
Fats - One of the three main
classes of foods and a source
of calories or energy in foods.
Fats help the body use some
vitamins and keep the skin
healthy.
Fiber - A substance found in
foods that comes from plants.
Fiber may help in the
digestive process and is
thought to lower cholesterol
and help control blood
glucose. Fiber grams are not
absorbed, and can be
subtracted from the Total
Carbohydrate amount for
foods or meals with more
than 5 grams of fiber.
Fingerstick - One method used
to test blood glucose by
taking a sample of blood for
testing from a finger using a
lancet or automatic finger
puncture device.
Fixed prime - Fills the cannula
with insulin. This is done after
you insert the infusion set
into your body and remove
the introducer needle.
Food bolus - A dose of insulin
given to cover the expected
rise in blood glucose that
occurs after eating.
Glossary
231
Glossary
G
Gastroparesis - A condition of
the digestive system that
slows down the emptying of
food from the stomach.
Glucagon - A hormone that
raises the level of glucose in
the blood. The alpha cells of
the pancreas make glucagon
when the body needs to put
more sugar into the blood.
Glucagon primarily causes
sugar stored in the liver to be
released into the blood
stream.
Glucagon Emergency Kit - An
injectable form of glucagon
that can be used to treat
severe low blood glucose
levels (hypoglycemia). A
Glucagon Emergency Kit can
be obtained in a pharmacy
with a prescription.
Glucose tablet - A simple sugar
in tablet form, that is taken
orally to treat low blood
glucose (hypoglycemia).
Glucose tablets can be
purchased from the Medtronic
Diabetes online store at
www.medtronicdiabetes.com
as well as drug stores and
pharmacies.
H
HbA1c - Glycosylated
hemoglobin
HDL - High-density lipoprotein A
complex of lipids and proteins
in approximately equal
amounts that functions as a
transporter of cholesterol in
the blood.
High Glucose - The pump alerts
if the sensor indicates that
your sensor glucose
measurement is at or above
this value. You have the
option to turn this feature on
or off.
High Snooze - For the High
Glucose alert, allows you to
set the delay between the
first alert and any subsequent
alerts. This allows you to
avoid getting an alert every
five minutes until the
condition is corrected.
Hold - Press and continue to
press a pump button.
Hyperglycemia - A condition in
which blood glucose values
are elevated above target.
Symptoms may include
excessive thirst, frequent
urination, dry mouth,
headache, fatigue, and
blurred vision.
Hypoglycemia - A condition in
which blood glucose values
are low below 3.9 mmol/L (70
mg/dL). Symptoms may
include behavioral changes,
pale complexion, hunger,
sweating, sudden weakness,
headache, confusion,
drowsiness, non-
responsiveness to questions,
and the most serious: seizure,
convulsion or loss of
consciousness.
I
Idle - The pump is at the HOME
screen.
Increment - A small increase or
decrease in a measurement.
Indication - A condition that
makes a specific medical
treatment or procedure
advisable or necessary.
Infusion set - Flexible tubing
with a reservoir connector
and an infusion site. This
tubing delivers insulin from
the pump to the body.
Infusion site - The end of the
infusion set held to the body
with a tape. It consists of a
cannula and an introducer
needle.
Insulin - A hormone that helps
the body use glucose (sugar)
for energy. The beta cells of
the pancreas make the
insulin.
Insulin sensitivity - The amount
by which your blood glucose
(BG) level is reduced by one
unit of insulin. (Bolus Wizard
feature data)
Insulin to carbohydrate ratio -
The amount of insulin
required to cover a given
number of carbohydrates.
This ratio helps you estimate
Glossary
232
what size bolus you should
take when eating
carbohydrates. Your
healthcare professional will
determine your insulin to
carbohydrate ratio.
Introducer needle
(1) This needle allows the
insertion of a cannula or a
sensor into the subcutaneous
tissue. It is removed and
discarded after insertion
leaving only the cannula or
the sensor in the body.
(2) A chemical created by the
human body when there is not
enough insulin in the blood.
IPX7 - Protected against water
immersion. Immersion for 30
minutes at a depth of up to 1
meter.
K
Ketone
kPa (kilopascal) - A unit of
measurement for quantifying
force. Used to measure
atmospheric pressure.
Equivalent to 10,000 dynes
per square centimeter.
L
LDL (Low-density lipoprotein) -
A complex of lipids and
proteins, with greater
amounts of lipid than protein,
that transports cholesterol in
the blood.
Link - To turn on and setup the
meter option that allows the
pump to receive BG readings
from a meter that
communicates with your
pump.
Log - A record of measurements,
alarms or other actions by
your pump.
Low Glucose - The pump alerts
if the sensor indicates that
your sensor glucose
measurement is at or below
this value. You have the
option to turn this feature on
or off.
Low resv warning -
Programmable warning that
notifies you with an alert
when either a specified
number of units remain in the
reservoir or a specified
amount of time remains
before the reservoir will be
empty.
Low Snooze - For the Low
Glucose alert, allows you to
set the delay between the
first alert and any subsequent
alerts. This allows you to
avoid getting an alert every
five minutes until the
condition is corrected.
M
Manual bolus - Selectable item
available in the BOLUS MENU
when the Bolus Wizard
feature is active. It is one
method of programming a
bolus without the Bolus
Wizard feature. (see Set
bolus)
Manual prime - Manual prime
fills the infusion set tubing
with insulin before you insert
the set into the body. Manual
prime is only available after
you rewind your pump.
Max basal rate - The maximum
amount of basal insulin that
the pump will deliver at one
time. (set by the user)
Max bolus - The maximum
amount of bolus insulin that
the pump will deliver at one
time. (set by the user)
Meter - An optional blood
glucose meter powered by
MWT1 technology. Your pump
can be programmed to
receive your BG readings from
this meter.
Meter option - Feature that
allows the pump to receive
BG readings from a meter
powered by MWT1 technology.
Mineral - Minerals and vitamins
are essential micronutrients,
needed in very small amounts
to keep our bodies functioning
properly.
Missed Data - The pump alerts if
it has not received data from
the sensor for an amount of
time that you set.
Glossary
233
Glossary
MWT1 technology - MWT1 is the
wireless Radio Frequency (RF)
technology that is used to
transmit information from the
meter to the pump. You can
program your pump to
automatically receive your BG
reading from this meter.
N
Normal bolus - An immediate
delivery of a specified unit
amount of insulin.
Normal mode - Regular
operating mode. No special
features are active, no alert
or alarm conditions exist.
Insulin delivery is normal
during this mode.
Now portion - The normal
portion of a Dual Wave bolus.
The now portion delivers
immediately and is then
followed by the Square
portion.
Nutrition - The process by which
the body draws nutrients from
food and uses them to make
or mend cells.
Nutrition label - Used to
determine the nutritional
content of foods, as well as
calories, carbohydrates, fats,
proteins, vitamin and mineral
content.
Nutritional content - The
breakdown of nutrients found
in a particular food.
O
Occlusive dressing - A bandage
that seals a wound from air or
bacteria.
P
Pattern A/B - Basal pattern that
supports activities that are
not a part of your day-to-day
routine, but are normal in
your lifestyle. Such activities
could be a sport that you do
once a week or a change in
your sleep pattern over the
weekend, extended periods of
higher or lower activity, or
menses.
Post-prandial reading - Blood
glucose reading taken after a
meal
Press - To push and release a
button.
Prime - To fill (prime) the
infusion set with insulin. See
Fixed prime, and Manual
prime.
Protein - One of the three main
sources of energy or calories
in food. Proteins are made of
amino acids, which are called
the building blocks of the
cells. The cells need proteins
to grow and to mend
themselves. Protein is found
in many foods such as meat,
fish, poultry, and eggs.
PSI - Pound-force per square
inch.
Pump S/N - Pump S/N is the
serial number of the pump
currently in use.
R
Reservoir - The syringe that
holds insulin.
Reset - To set again, or change
the information (reset your
glucose values).
Resume - Restarts basal delivery
after the pump is suspended.
Rewind - The pump drive moves
back to its starting position to
prepare the pump for a new
reservoir.
RF - Radio frequency
S
Scroll - Press the up or down
arrow buttons to move
through the screen text.
Select - Pressing the up or down
arrow buttons to highlight a
desired screen item.
Sen-serter - The Sen-serter is
indicated as an aid for
insertion of the Medtronic
Diabetes glucose sensor.
Sensitive - Easily irritated
(sensitive skin), or able to
measure very slight
differences (a sensitive
instrument).
Glossary
234
Sensor Age - Sensor age is the
amount of time, in days and
hours, since the sensor was
first inserted.
Sensor features - Optional
continuous glucose monitoring
capabilities.
Sensor glucose (SG) - Glucose
levels in the fluid under your
skin measured by the sensor.
Set - To enter or establish a
value for your pump (set your
BG Reminder).
Set bolus - Selectable item
available in the BOLUS MENU
when the Bolus Wizard
feature is inactive. One
method of programming a
bolus without the Bolus
Wizard feature. (See Manual
bolus.)
S/N# - Serial number.
Sodium - Is an essential element
for your body. As table salt, it
is used freely in foods as a
flavor enhancer and a
preservative. Sodium is listed
on food labels and is
expressed in milligrams.
Special mode - Operating mode
that indicates one or more
special functions is active or a
condition exists that requires
attention.
Square Wave
®
bolus -
Immediate bolus delivered
evenly over a specified time
period (30 minutes to 8
hours).
Square Wave portion - (Sq) The
second part of a Dual Wave
bolus. The Square Wave
portion delivers evenly over a
period of time after the NOW
portion delivers.
Standard pattern - Your normal
basal that supports your usual
day-to-day activity. When the
Patterns feature is off, the
pump uses your standard
basal pattern.
Status - The condition of a part
of your system (battery
status, alarm/alert status).
Status screen - Displays the
current operations of the
pump, including active
functions, the most recent
basal and bolus deliveries,
reservoir information, and
battery condition.
Step - Measurement of insulin
that you set and use for Easy
Bolus and other bolus
deliveries.
Submerge - To place under
water or cover completely
with water.
Suspend - Function that stops all
current insulin deliveries. The
basal delivery is paused until
restarted.
T
Temp - Temporary
Temp basal - (Tmp basal)
Temporary one-time basal
insulin with a specified
amount and duration. Used to
support insulin needs for
special activities or conditions
that are not part of the
normal daily routine.
Timeout - If you do not enter
information or do something
in a specific length of time,
your pump automatically
stops what it is doing and
goes to the HOME screen.
Transmtr ID - The serial number
of the transmitter currently in
use.
Glossary
235
Glossary
Glossary
236
Index
Numerics
3-hour graph 128
24-hour graph 129
A
accessories 3
activity guard 4
CareLink USB 3
holster 4
leather case 4
meter 3
pump clip 4
remote control 3
sensor 3
transmitter 3
active insulin 81
Active insulin time
setting 82
adverse reactions 10
alarm clock 141
Alarm conditions 165
alarm frequency 177
ALARM HISTORY screen 135
Alarm Snooze
default settings 196
setting 114
Alarm types
A (alarm) 165
Auto off 165
Batt out limit 165
Bolus stopped 165
Button error 166
check settings 155
Check settings 166
E (error) 166
Empty reservoir 166
Failed batt test 166
Max delivery 166
Motor error 166
no delivery 153
No delivery 167
No reservoir 167
Off No power 167
Reset 167
Weak battery 167
Alarms
responding to 164
review 135
alert frequency 177
alert type
Setting your 135
alerts 160
Bad Sensor 162
Bad Transmtr 162
Cal Error 162
Charge Transmtr 162
for alert conditions 160
low reservoir alert 160
High XXX MG/DL 163
Lost Sensor 161
Low Transmtr 161
Low XX MG/DL 163
Meter BG By 163
Meter BG Now 163
sensor 160
Sensor End 162
Sensor Error 163
Weak Signal 160
areas
inserting infusion set 65
inserting sensor 119
Attention mode 34
auto-off 136
B
backlight 33, 229
Bad Sensor alert 162
bad transmitter battery 170
Bad Transmtr alert 162
BASAL MENU 35
basal rate
daily 52
maximum 54
patterns 100
programming 51
stop delivery of 55
temp 102
bathing 133
battery, charger
low 170
Index
237
Index
battery, pump 173, 183
install 28
removing 154
short life 154
beep/vibrate, alert type 33
BG Reminder 41, 49
BG Targets 80
BG units 78
default settings 196
selecting 112
bleeding 8, 10, 121
Block feature 143
Turn on 143
BLOCK OPTION screen 143
blood glucose
checking 123
entering incorrect
number 162
rapid change 162
blood glucose units
default settings 196
bolus
basic 41
details 47
history 46
manual 92
maximum limit 48
stop delivery of 55
bolus delivery
Bolus Wizard 85
Easy Bolus 97
bolus delivery, methods for
normal
remote control 3
BOLUS MENU 35
bolus types
Dual Wave 91
Normal 41
Square Wave 91
Bolus Wizard
about 71
entering meter BG 124
estimate details 47
HIGH BG warning 75
how it works 75
LOW BG warning 76
MAX BOLUS EXCEEDED
warning 76
maximum delivery 76
on/off 77
review settings 82
settings 72, 73, 74, 75
Active insulin time 75
BG Ttarget range 74
BG units 73
Carb ratios 73
Carb units 73
Insulin sensitivity 74
specifications 186
your BG reading 71
bruising 10
buttons, pump 27, 29
C
Cal Error alert 162
Cal Reminder 163
default settings 196
setting 115
CAL REMINDER screen 115
CALIBRATE HISTORY 131
calibrating
frequency 123
calibrating sensor 123
frequency 123
calibration
successful 131
values 131
when to 123
Calibration history 131
Carb ratios
exchanges 78
grams 78
carb units 77
carbohydrate counting 22
CareLink Personal Software 149
CareLink USB 3
Charge Transmtr alert 162
charger 3
battery power 117
connecting with
transmitter 118
low AAA battery 170
transmitter 117
water-tightness 174
charging
transmitter 117
CHECK BG message 49
checking
blood glucose 123
choosing insertion site 120
cleaning
Sen-serter® 175
transmitter 174
cleaning the pump 174
Clear settings 147
connecting transmitter to
charger 118
sensor 122
tester 167
consumables 3
correction bolus 41, 85, 86
D
Daily totals 138
DAILY TOTALS screen
Review 137
default settings
Alarm Snooze 196
BG Units 196
Cal Reminder 196
High Glucose Alert 196
High Snooze 196
insulin delivery 193
Low Glucose Alert 196
Low Snooze 196
Index
238
Missed Data 196
sensor 196
sensor features 196
discarding
transmitter 174
disconnecting
sensor from transmitter 132
tester from transmitter 168
transmitter from sensor 132
disposing
introducer needle 121
downloading pump data 3
Dual Wave bolus programming
with Bolus Wizard 95
without Bolus Wizard 92
dual/square option
on-off 92
E
Easy bolus 97
cancel 99
deliver 98
step value 97
electromagnetic emissions 197
Electromagnetic
environment 203
electromagnetic immunity 198
emergency kit 2
emissions 197
entering
meter BG 123, 124
alert 123
Bolus Wizard 124
frequency 123
manually 124
ESC 5
cancels programming 36
Examples of real-time sensor
glucose graphs 128
exercising
precaution 10
express bolus button 30
F
filling
reservoir 58
FIND LOST SENSOR 169
fixed prime 67
food bolus 41, 73, 75, 85, 86
food entry 72, 85
food label 23
frequency
alarm 177
alert 177
G
graphs 127
differentiating between
current and earlier
events in 126
opening 126, 127
real-time glucose 125
real-time sensor glucose
levels 127
green light
transmitter 122, 168
H
high BG levels
Bolus Wizard and 75
HIGH BG warning 75
High Glucose Alert
default settings 196
High Glucose limit
setting 112
High Glucose Limit
alert 163
High sensor glucose alert 163
High Snooze
default settings 196
setting 113
High Snooze alert
setting 113
history
alarms 135
bolus delivery 45
calibration 131
sensor alert 131
user settings 147
HOME screen 31
I
infection 8, 10
infusion set 3, 57
changing your 60
inserting 65
best sites 65
rotating sites 65
initialization 123
inserting sensor
best sites 119
guidelines 120
procedure 120
Sen-serter 120
sites to avoid 120
steps before 119
stopping bleeding 121
warning 119
Insulin sensitivity 79
interference
transmitter 8
introducer needle 120
disposing 121
removing 121
irritation
occlusive dressing 8, 10
skin 8
ISIG 168
I.V. Prep 120
Index
239
Index
L
Language setting 40
list
calibration values 131
sensor alert 131
Lock keypad feature 144
Lost Sensor alert 161
low battery
affects meter option 84
affects remote control 142
affects selftest 144
affects your alert type 135
low BG levels
Bolus Wizard and 76
LOW BG warning 76
Low Glucose
setting alert frequency 114
Low Glucose Alert
default settings 196
Low Glucose limit
setting 113
Low Glucose Limit
setting 163
low reservoir alert 160
Low resv warning feature 136
Low sensor glucose alert 163
Low Snooze
default settings 196
setting 114
low transmitter battery 161
Low Transmtr alert 161
LR-03 battery
low 170
M
MAIN MENU 35
manual prime 64
manually entering meter BG 124
Max basal rate 54
MAX BOLUS EXCEEDED
warning 76
maximum bolus limit 48
menus 35
meter 3, 71
entering BG 123, 124
Meter BG By alert 163
Meter BG Now alert 163
METER BG NOW alert
setting frequency 114
meter IDs
add, delete, review 84
meter option 83
rules 84
METER OPTION screen 84
MiniLink transmitter 3
Missed Data
default settings 196
Missed Data feature 116
description 116
modes
Attention 34
Normal 34
Special 34
MWT1 3
MWT1 technology 71, 124
N
needle guard 120
no delivery 153
Normal bolus
with Bolus Wizard 85
without Bolus Wizard 41
Normal mode 34
Now 96
Now portion
of the Dual Wave bolus 93, 96
nutrition 22
O
occlusive dressing
irritation 8, 10
reaction 8, 10
P
pain 10
Patterns
basal 100
on/off 100
pattern A, B 100
Program a basal pattern 100
Select a basal pattern 101
standard basal pattern 100
Personal reminders 141
personal settings 72
practicing 62
precaution
sensor 10
preparing sensor sites 120
prime
fixed 67
manual 64
PRIME MENU 36
procedure, inserting sensor 120
programming sensor 111
pump
battery compartment 27
buttons 27, 29
if you remove your 37
screen 27, 31
using in water 158
warranty 206
R
reaction to occlusive
dressing 8, 10
Index
240
reading the real-time glucose
graphs 125
real-time glucose graphs
reading 125
real-time sensor glucose
levels 127
RECONNECT OLD SENSOR 169
Reminder, alarm clock 141
remote control 3
remote control IDs
add, delete, review 142
Remote control option
turn on 142
removing
introducer needle 121
needle guard 120
sensor 132
removing your pump 37
reservoir
filling the 58
inserting in your pump 62
low reservoir warning 136
plunger rod 58
removing the 60
transfer guard 58
tubing connector 27
reservoirs 3
Restore settings 146
REVIEW SETTINGS screen 116
reviewing
sensor settings 116
rewinding your pump 62
during practice 62
RF communications 203
RF features
meter 83
remote control 141
use on aircraft 10
S
Save settings 145
scroll bar 33
Selftest 144, 157
Sen-serter
inserting sensor 119, 120
removing from sensor 121
Sen-serter®
cleaning 175
sensor
alerts 131, 160
history 131
list 131
best sites 119
calibrating 123, 163
caution inserting 119
choosing insertion site 119
connecting with
transmitter 122
default settings 196
disconnected 161
disconnecting from
transmitter 132
improper insertion 119
information 125
inserting 119, 120, 121
bleeding 121
guidelines 120
procedure 120
ISIG 168
life 162
precaution 10
programming 111
reconnecting 169
removing 132
setting up 111
signal 119
signals 163
too high 163
too low 163
sites 120
preparing 120
stabilizing 162
starting 119
status screens 125
Transmitter ID 115
using in water 133
warnings 8, 119
SENSOR ALARM HISTORY 131
sensor alerts
Bad Sensor 162
Bad Transmtr 162
Cal Error 162
Charge Transmtr 162
High XXX MG/DL 163
Lost Sensor 161
Low Transmtr 161
Low XX MG/DL 163
Meter BG By 163
Meter BG Now 163
Sensor End 162
Sensor Error 163
Weak signal 160
Sensor End alert 162
Sensor Error alert 163
Sensor function 35
sensor glucose levels
graphs 127
SENSOR ON/OFF screen 111
SENSOR STATUS 125
Sensor update history 131
SET ALARM SNOOZE screen 114
SET CAL REMINDER screen 115
SET MISSED DATA screen 116
SET TRANSMITTER ID screen 115
setting
Alarm Snooze 114
Cal Reminder 115
High Glucose limit 112
High Snooze 113
Low Glucose limit 113
Low Snooze 114
setting up sensor 111
settings, reviewing 116
showering 133
sites
inserting sensor 119
Index
241
Index
skin
bleeding 8, 10
bruising 10
infection 8, 10
redness 10
swelling 8, 10
Special mode 34
Square portion
of the Dual Wave bolus 96
Square Wave bolus programming
with Bolus Wizard 95
without Bolus Wizard 92
starting sensor
procedure 123
steps before 119
STATUS screen 36
Step value setup 97
stopping bleeding 121
supplies, ordering 4
Suspend function 35, 55
swelling 8, 10
swimming 133
system initialization 123
T
Temp basal
Canceling a 108
Deliver a 107
rates 102
Selecting type 107
verify delivery 108
temp basal types
Insulin rate 103
Percent of basal 105
tenderness 10
terms and symbols 5
test plug. See tester 168
tester 3
connecting to
transmitter 167
disconnecting from
transmitter 168
transmitter 117
water-tightness 174
therapy follow-up
Recommended 151
time and date, setting 39
transmitter
bad battery 170
battery 117
before attaching 121
charger 117
charging 117
charging complete 118
cleaning 174
connecting with charger 118
connecting with sensor 122
depleted battery 162
discarding 174
disconnected 161
disconnecting from
sensor 132
end of life 170, 171
green light 122, 168
interference 8
light 122
lights 118
low battery 161
signal 119
tester 3, 117
troubleshooting 117
using in water 133
warning 8
Transmitter ID
setting 115
troubleshooting 169
U
user safety
adverse reactions 10
User settings 145
User Settings screen 158
Utilities 135
UTILITIES MENU 36
V
viewing
calibration values 131
sensor alarm history 131
W
warnings
sensor 8
transmitter 8
warranty 206
water, using your system in 133
water-tightness 133
charger 174
tester 174
Weak Signal 160
Weak Signal alert 160
Index
242
6025250-011C 112106
Dostları ilə paylaş: |