required. For a subject with weak pulse due to shock,
25
will be more sensitive to interference.
2. For those with a substantial amount of staining dilu-
tion drug (such as methylene blue, indigo green and
acid indigo blue), or carbon monoxide hemoglobin
(COHb), or methionine (Me+Hb) or thiosalicylic he-
moglobin, and some with icterus problem, the SpO
2
determination by this monitor may be inaccurate.
3. The drugs like dopamine, procaine, prilocaine, li-
docaine and butacaine may also be a major factor
blamed for serious error of SpO
2
measure.
4. As the SpO
2
value serves as a reference value for
judgement of anemic anoxia and toxic anoxia, some
patients with serious anemia may also report good
SpO
2
measurement.