6.3 Visual inspection
In the visual inspection, it is checked whether the lux metre submitted for validation (including software version) conforms to the approved type. Special attention must be paid to checking the correctness of the markings as defined in Article 4.1.
The lux metre is also inspected to determine whether or not it is mechanically damaged. Lux metres with an electronic display are checked to determine whether all characters on the display are visible after the device is connected to the mains power supply.
Lux metres which are not in conformity with the approved type are not tested further, nor are lux metres which are damaged.
6.4 Preliminary function test
The preliminary function test is conducted in accordance with Article 5.3.
6.5 Accuracy test
The accuracy test may be tested either:

  • using a set of reference photometric illumination lamps with defined colour temperature values pursuant to Article 5.4.2, or

  • using a set of reference photometric illumination lamps with defined colour temperature values pursuant to Article 5.4.3.

6.6 Linearity test
The test linearity on the lux metre being validated is conducted in accordance with Article 5.5.
6.7 Determination of the correction coefficients for types of light sources other than standardised source A
The spectral adaptation of a lux metre’s spectral responsivity to the relative spectral luminous efficiency for photopic vision V (λ) is verified by comparing the lux metre being evaluated to the reference on working sources other than CIE standardised source A.
Working light sources used to verify the spectral adaptation of a lux metre’s spectral responsivity to the relative spectral luminous efficiency for photopic vision V (λ):

  • sodium vapour lamp light source,

  • a light source similar to standardised source D65 (with a colour temperature of roughly 6 500 K), which, in spectral terms, is akin to daylight,

  • light source of a white fluorescent lamp with luminescent material (with a colour temperature of roughly 2 800 K),

  • light source of a white fluorescent lamp with luminescent material (with a colour temperature of roughly 5 500 K).

Correction coefficients are determined for one luminance value of a photometric lamp; the value is selected based on the range of the lux metre being validated. The correction coefficient value kj for the type of light j is established according to relation (10):
(10)
in which
is the value of the illuminance measured with the reference lux metre for illumination from source A,
 is the value of the illuminance measured with the reference lux metre for illumination from source j,
 is the value of the illuminance measured with the reference lux metre for illumination from source A,
The values of the correction coefficients kj for each individual type of light source j are specified on the validation sheet.