conformity of the metrological and technical characteristics specified by the manufacturer in the documentation with the requirements of this regulation, specified in Articles 2, 3 and 4;
the completeness and condition of the lux metre according to the specified technical documentation,
conformity of the lux metre’s software version with the version specified by the manufacturer.
5.3 Preliminary function test The device must be placed in a temperature stabilised laboratory (at 25 ± 2) °C for a sufficient length of time for its temperature to adjust before measuring commences.
The photoelectric transducer is mounted firmly to a holder at the optical zero of the photometric bench and the display unit is placed in the prescribed position.
A cursory check of the lux metre’s functioning is conducted in an undarkened photometric laboratory by measuring a light background in the laboratory environment.
After the photometric laboratory has been darkened, the device’s zero deviation is checked (and set, if necessary), if the lux metre so allows.