5 Type approval of the measuring instrument
5.1 General
The type approval process is made up of the following tests:

  • external inspection,

  • preliminary function test,

  • accuracy test,

  • linearity test,

  • relative spectral responsivity test,

  • test of sensitivity to UV and infrared radiation,

  • temperature dependence test,

  • directional dependence test,

  • test of resistance to external effects:

  • test of resistance to boundary temperatures for storage and transportation,

  • test of protection from water seepage and infiltration by foreign particles (degree of protection offered by the cover);

  • electromagnetic compatibility tests:

  • test of resistance to radiated high-frequency electromagnetic fields,

  • test of resistance to electrostatic discharge.

5.2 External inspection
The external inspection of the lux metre assesses the following:

  • the completeness of the required technical documentation,

  • conformity of the metrological and technical characteristics specified by the manufacturer in the documentation with the requirements of this regulation, specified in Articles 2, 3 and 4;

  • the completeness and condition of the lux metre according to the specified technical documentation,

  • conformity of the lux metre’s software version with the version specified by the manufacturer.

5.3 Preliminary function test
The device must be placed in a temperature stabilised laboratory (at 25 ± 2) °C for a sufficient length of time for its temperature to adjust before measuring commences.
The photoelectric transducer is mounted firmly to a holder at the optical zero of the photometric bench and the display unit is placed in the prescribed position.
A cursory check of the lux metre’s functioning is conducted in an undarkened photometric laboratory by measuring a light background in the laboratory environment.
After the photometric laboratory has been darkened, the device’s zero deviation is checked (and set, if necessary), if the lux metre so allows.