47
5.4.
Results of the fluorescence studies
The fluorescent light emitted by the plastic scintillators upon exc itation with a
229 nm laser was measured. The photo-bleaching of
this fluorescent light was
minimised by reducing the laser exposure time as well as scanning the laser over
the region being measured. These conditions coupled with the detector
positioning resulted in fluorescence off the surface region being measu red rather
than observing “bulk” effects. Photo-bleaching time correlation tests were also
performed in order to correct the data.
The average spectra of three measurements for the un -irradiated
and irradiated
regions respectively for all the samples unde r testing are shown in Appendix D.
Their corresponding photo-bleaching vs time correlation curves are also shown
alongside each spectra.
5.4.1.
Fluorescence peak features
Several fluorescence peak features can be observed amongst the spectra. Peaks
falling within the wavelength region of 300 -375 nm correlate to fluorescence off
the PVT/PS base. Since this fluorescence is predominantly from the
benzene ring
structure in both base types, the same “two-peak” feature over this region is seen
in the un-irradiated spectra for all the different scintillator types tested.
In the wavelength region of 375-500 nm, the fluorescence correlates to that of
the fluor dopants. These peak features therefore
vary between the different
scintillator types. The peaks reach a maximum between 400 -430 nm for all the
blue emitting scintillators.
EJ260 shows additional peak features at ~460 nm and 490 nm, however these are
weaker than expected. It is possible that due to the experimental conditions, the
emitted light measured had less spatial interaction with the fluors responsible for
shifting the emitted light into the green wavelength region.
48
The effect of limited spatial interaction could also explain why peaks in the base
fluorescence region have a high intensity comparable to the fluor regions for both
the green and blue emitting scintillators. If fluorescence was measured over the
bulk rather than at the surface, majority of the “base emitted light” would be re-
absorbed and shifted to higher wavelengths through interaction with the fluors.
For the samples exposed to radiation damage, a general trend in the features are
observed. For doses of ~0.8-1 MGy, an overall intensity loss occurs, with a more
distinct loss to fluorescence in the fluor emission regions (375 -500 nm). At this
dose, very minimal transmission losses were observed
in the transmission
spectra, although a feature indicating loss to absorption by the fluors occurred.
At progressively higher doses, the fluorescence
intensity is further decreased,
with significant loss over both the base and fluor regions. At the 80 MGy dose
range, a very weak signal is obtained with an additional peak feature appearing
in some of the spectra. This peak feature could not be correlated to any particular
damage effect and may be influenced by several factors such as sample thickness
and dose rate.
Dostları ilə paylaş: