Cost Effective Quantification of Trace Level edta (Ethylene Di-amine Tetra Acetic Acid) by Titrimetry in Active Pharmaceutical Ingredients


Table 5: Method precision results of EDTA (0.05%) spiked in 6-APA. No



Yüklə 0,55 Mb.
Pdf görüntüsü
səhifə9/11
tarix24.12.2022
ölçüsü0,55 Mb.
#77727
1   2   3   4   5   6   7   8   9   10   11
cost-effective-quantification-of-trace-level-edta-ethylene-diamine-tetraacetic-acid-by-titrimetry-in-active-pharmaceutical-ingredi

Table 5: Method precision results of EDTA (0.05%) spiked in 6-APA.
No.
EDTA Content(%w/w)
1
0.0603
2
0.0500
3
0.0532
4
0.0560
5
0.0533
6
0.0480
Mean
0.0535
SD^
0.004
%RSD*
8.22
^ Standard deviation
* Relative standard deviation
Accuracy
Accuracy of the method was performed by recovery experiment using standard addition method technique. The 
recoveries of I, II and III were determined by spiking EDTA at three different levels ranging from 0.05% to 0.15% 
into drug substance. These samples were prepared as per the procedure and analysed in triplicate and the percentage 
recoveries were calculated. The recovery value of EDTA Ranged from 98 % to 100.5%.The average recovery of three 
levels (Nine determinations).The accepted limits of recovery are 98 % to 100.5% and all observed date are within 
the required range which indicates good recovery values and hence the accuracy of the method developed (Table 6).
Table 6: Recovery results from spiking of 6-APA with EDTA.
Accuracy 
(Average of triplicates)
Level-I (LOQ)
Level-II (100%)
Level-III (150%)
Amount Added (%)
0.0535
0.1064
0.1554
Amount Found (%)
0.0535
0.1044
0.1554
Recovery (%)
100.0
98.09
100.0
RSD (%)
1.35
0.31
1.00
Robustness
The Robustness of an analytical method is found to be robust by performing small variations in the analytical method 
parameters it will not affect the results. Based on this study it is concluded that the method is Robust.
Sensitivity
The sensitivity of measurement of EDTA content (standard) (Figure 7) by use of the proposed method was estimated 
in terms of the limit of quantitation (LOQ), Limit of detection (LOD). The limit of detection (LOD) (Figure 8) and 
Limit of quantitation (LOQ) (Figure 9) were found to be 0.0160% and 0.0350% respectively. Optical characteristics 