PREPARATION The reagent is ready to use.
STORAGE AND STABILITY All the components of the kit are stable until the expiration date on the
label when stored tightly closed at 2-8ºC, protected from light and
contaminations prevented during their use.
Do not use reagents over the expiration date.
SIGNS OF REAGENT DETERIORATION: -
Presence of particles and turbidity.
-
Blank absorbance (A) at 505 nm
0,32
ADDITIONAL EQUIPMENT -
SPIN640 / SPIN640Plus Autoanalyzer.
-
General laboratory equipment.
SAMPLES Serum or plasma, free of hemolysis
1
:
Serum should be removed from the clot as quickly as possible.
Stability of the sample: Glucose in serum or plasma is stable at 2-8º
for 3 days.
QUALITY CONTROL Control sera and calibrators are recommended to monitor the
performance of assay procedures: SPINTROL H Calibrator,
SPINTROL H Normal and Pathologic (Ref. 1002011, 1002120 and
1002210).
If control values are found outside the defined range, check the
instrument, reagents and technique for problems.
Each laboratory should establish its own Quality Control scheme and
corrective actions if controls do not meet the acceptable tolerances.
BARCODED REAGENTS LOAD MUST BE PRECEDED OF A SPINREACT “DATABASE” COPY INTO THE ANALYZER SOFTWARE. IT IS AVAILABLE UNDER REQUEST TO SPINREACT. SPIN640 APPLICATION
SPIN640Plus APPLICATION
The Calibration is stable until 36 days. After this period the Calibration must be performed again in order to obtain good results.